TR 60 - TR 1050

With their maximum working temperature of 300 °C and forced air circulation, the ovens achieve a perfect temperature uniformity which is much better than in ovens of most competitors. They can be used for various applications such as e.g. drying, sterilizing or warm storing. Ample warehousing of standard models provides for short delivery times.

  • Tmax 300  °C
  • Working temperature range: + 5  °C above room temperature up to 300  °C
  • Ovens TR 60 - TR 240 designed as tabletop models
  • Ovens TR 450 and TR 1050 designed as floor standing models
  • Horizontal, forced air circulation results in temperature uniformity better than +/- 5 °C
  • Stainless steel chamber, alloy 304 (AISI)/(DIN material no. 1.4301), rust-resistant and easy to clean
  • Large handle to open and close the door
  • Charging in multiple layers possible using removeable grids (number of removeable grids included, see table to the right)
  • Large, wide-opening swing door, hinged on the right with quick release for models TR 60 - TR 450
  • Double swing door with quick release for TR 1050
  • TR 1050 equipped transport rollers
  • Infinitely adjustable exhaust at the rear wall with operation from the front
  • PID microprocessor control with self-diagnosis system
  • Solid state relays provide for low-noise operation
  • Defined application within the constraints of the operating instructions
Additional Equipment
  • Over-temperature limiter with adjustable cutout temperature for thermal protection class 2 in accordance with EN 60519-2 as temperature limiter to protect the oven and load
  • Infinitely adjustable fan speed of the air circulation fan
  • Window for charge observing
  • Further removeable grids with rails
  • Side inlet
  • Stainless steel collecting pan to protect the furnace chamber
  • Safety Technology according to EN 1539 for charges containing liquid solvents (TR .. LS) up to model TR 240 LS, achievable temperature uniformity +/- 8 °C
  • Transport costors for model TR 450
  • Various modifications available for individual needs
  • Upgrading available to meet the quality requirements of AMS 2750 E or FDA
  • Process control and documentation via VCD software package for monitoring, documentation and control
 
Model Tmax Inner dimensions in mm Volume Outer dimensions in mm Heating Electrical Weight Grids in- Grids Max.
  °C w d h in l W D H power in kW2 connection in kg cluded max. total load
TR 60 300 450 390 350 60 700 610 710 3 1-phase 90 1 4 120
TR 60 LS 260 450 360 350 57 700 680 690 6 3-phase 92 1 4 120
TR 120 300 650 390 500 120 900 610 860 3 1-phase 120 2 7 150
TR 120 LS 260 650 360 500 117 900 680 840 6 3-phase 122 2 7 150
TR 240 300 750 550 600 240 1000 780 970 3 1-phase 165 2 8 150
TR 240 LS 260 750 530 600 235 1000 850 940 6 3-phase 167 2 8 150
TR 450 300 750 550 1100 450 1000 780 1470 6 3-phase 235 3 15 180
TR 1050 300 1200 670 1400 1050 1470 940 1920 9 3-phase 450 4 14 250

Please leave your enquiry here, we will reply as soon as possible.
Name*  
Company Name  
Product Interested  
Quantity  
Email*  
Contact No.*  
Ship to Name  
Shipping Address  
Attachment  
*only support gif, jpeg, jpg, png, pdf
Messages*  


You have 0 items in you cart. Would you like to checkout now?
0 items
Switch to Mobile Version
available in
Subscribe Newsletter