Muffle Furnaces with Embedded Heating Elements in the Ceramic Muffle
Previous | 5 / 8 | Next |
View old post here.
L, LT 9/11/SKM
L, LT 9/11/SKM
We particularly recommend the L 9/11/SKM model if your application involves aggressive substances. The furnace has a ceramic muffle with embedded heating from four sides. The furnace thus combines a very good temperature uniformity with excellent protection of the heating elements from aggressive atmospheres. Another aspect is the smooth, nearly particle free muffle (furnace door made of fibre insulation), an important quality feature for some ashing processes.
- Tmax 1100 °C
- Muffle heated from four sides
- Furnace chamber with embedded ceramic muffle, high resistance to aggressive gasses and vapours
- Dual shell housing made of sheets of textured stainless steel
- Only fiber materials are used which are not classified as carcinogenic according to TRGS 905, class 1 or 2
- Optional flap door (L) which can be used as work platform or lift door (LT) with hot surface facing away from the operator
- Adjustable working air inlet in the door
- Exhaust air outlet in rear wall of furnace
- Solid state relays provide for lownoise operation
- Defined application within the constraints of the operating instructions
- NTLog Basic for Nabertherm controller: recording of process data with USB-flash drive
Additional Equipment
- Chimney, chimney with fan or catalytic converter
- Over-temperature limiter with adjustable cutout temperature for thermal protection class 2 in accordance with EN 60519-2 as temperature limiter to protect the furnace and load
- Protective gas connection to purge with non-flammable protective or reaction gases
- Manual or automation gas supply system
- Observation hole in the door
- Process control and documentation via VCD software package for monitoring, documentation and control
Model | Tmax °C | Inner dimensions in mm | Volume in l | Outer dimensions in mm | Connected load kW | Electrical connection | Weight in kg | Minutes to Tmax² | ||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
W | D | H | W | D | H | |||||||
L 9/11/SKM | 1100 | 230 | 240 | 170 | 9 | 490 | 505 | 580 | 3.4 | single-phase | 50 | 90 |
LT 9/11/SKM | 1100 | 230 | 240 | 170 | 9 | 490 | 505 | 580+320 | 3.4 | single-phase | 50 | 90 |